Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Philips PM 3320A: (1 Year) CAL VER /5500+SC300,8842 DATE: 03-Oct-97 AUTHOR: Fluke REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 147 NUMBER OF LINES: 881 CONFIGURATION: Fluke 5500A (SC) CONFIGURATION: Fluke 8842A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP This procedure was "cloned" from a CG5011 based 1.001 DISP procedure that was tested with the UUT. However, this 1.001 DISP procedure has not been tested with the UUT. 1.002 ASK+ D N 1.003 ASK- R P J F W 1.004 HEAD 4.1 GENERAL INFORMATION 1.005 DISP [32] WARNING 1.005 DISP Before switching-on, ensure that the instrument has been 1.005 DISP installed in accordance with Installation Instructions 1.005 DISP outlined in chapter 3.0 of the Operating Manual. 1.006 DISP This procedure is intended to: 1.006 DISP Check the instrument's specification. 1.006 DISP Be used for incoming inspection to determine the 1.006 DISP acceptability of newly purchased instruments and/or 1.006 DISP recently recalibrated instrument. 1.006 DISP Check the necessity of recalibration after the specified 1.006 DISP recalibration intervals. 1.007 DISP NOTE: 1.007 DISP The procedure does not check every facet of the 1.007 DISP instruments calibration; rather, it is concerned 1.007 DISP primarily with those parts of the instrument which are 1.007 DISP essential to measurement accuracy and correct operation. 1.007 DISP Removing the instruments covers is not necessary to 1.007 DISP perform this procedure. All checks are make from the 1.007 DISP outside of the instrument. 1.008 DISP If the test is started within a short period after 1.008 DISP switching on, bear in mind that steps may be out of 1.008 DISP specification, due to insufficient warming-up time. 1.008 DISP 1.008 DISP Warming-up time under average conditions is 30 minutes. 1.009 DISP The performance checks are make with a stable, well- 1.009 DISP focussed, low-intensity display. Unless otherwise noted, 1.009 DISP adjust the intensity and trigger-level controls as 1.009 DISP needed. 1.010 DISP NOTE: 1.010 DISP At the start of every check, the controls always occupy 1.010 DISP the preliminary settings AUTO SET position, unless 1.010 DISP otherwise stated. 1.011 DISP The input voltage has to be supplied to the A-input; 1.011 DISP unless otherwise stated. 1.011 DISP Unless otherwise stated, the switches TIME/DIV, SHIFT A 1.011 DISP and B, X-POSITION, TRACE INTENS and TEXT INTENS must be 1.011 DISP put in such a position that a good read-out of the 1.011 DISP phenomena of interest is obtained. 1.012 HEAD 4.3 PRELIMINARY SETTINGS 1.013 DISP Switch-on the instrument (no input signal). 1.014 RSLT =Verification of LCD segments. 1.015 EVAL Did all LCD segments illuminate for approx. 1 sec? 2.001 DISP Press AUTO. 2.002 RSLT =Verification of SOFTSTART condition. 2.003 MESS After adjustment of TEXT INTENS and TRACE INTENS is the 2.004 EVAL readout information and a horizontal line visible on the CRT? 3.001 MESS 3.002 HEAD {B.1 TRACE DISTORTION} 3.003 DISP Press VERTICAL COUPLING. 3.003 DISP Press softkeys to select ch.A & ch.B GROUND. 3.003 DISP Press softkeys to select ch.A & ch.B 1 MOhm. 3.003 DISP Set VERTICAL SHIFT ch.A to center trace on screen. 3.004 EVAL Is trace exactly parallel with the horizontal graticule lines? 3.005 DISP Adjust front panel TRACE ROTATION 3.005 DISP so that the trace is exactly parallel with the 3.005 DISP horizontal graticule lines. 3.006 JMP 3.004 4.001 MESS Shift the trace upwards and downwards. 4.001 MESS Does the deviation from the straight line exceed 0.1 div 4.002 EVAL measured outside the central 8x10 div? [N] 5.001 MESS 5.002 DISP Press AUTO. 5.003 DISP Connect the 5500A Scope Output to ch.A. 5.004 5500 0.2 1.6Vpp 2kH SI N 2W 5.005 DISP Press VERTICAL MODE. 5.005 DISP Press softkey to select A and B. 5.005 DISP Set VERTICAL SHIFT ch.A & ch.B to center traces on the 5.005 DISP screen. 5.005 DISP Press DISPLAY. 5.005 DISP Press softkey A versus B>. 5.005 DISP Press softkey to select R0 ON. 5.006 EVAL Is a vertical line of 8 div. displayed? 6.001 5500 Vpp 2kH SI SC S 6.002 MESS Shift this line to the left and right by means of the 6.002 MESS X-POSITION control. 6.002 MESS Does the deviation from the straight line exceed 0.1 div 6.003 EVAL measured outside the central 8x10 div? [N] 7.001 5500 Vpp 2kH SI SC S 7.002 DISP Shift the vertical line in the horizontal mid of the 7.002 DISP screen. 7.003 MESS Does the angle between the vertical graticule lines and 7.004 EVAL the displayed line exceed 0.5 degrees? [N] 8.001 MESS 8.002 HEAD {C.1 INPUT COUPLING: ch.A} 8.003 5500 120mVpp 1kH SI SC S L 8.004 DISP Press AUTO. 8.005 DISP Press VERTICAL COUPLING. 8.005 DISP Press softkey to select ch.A DC. 8.005 DISP Press softkey to select ch.A 50 Ohm. 8.005 DISP Set VERTICAL AMPL/DIV ch.A to 20mV. 8.005 DISP Set HORIZONTAL TIME/DIV to 20ms. 8.005 DISP Press TRIGGER COUPLING. 8.005 DISP Press softkey twice to select DC. 8.006 MESS Set TRIGGER LEVEL for a triggered display. 8.007 5500 40 240mVpp 10H SI SC N L 8.008 DISP Press VERTICAL COUPLING. 8.008 DISP Press softkey to select ch.A AC. 8.009 EVAL Is the amplitude of the displayed signal >=4.3 div? 9.001 MESS 9.002 DISP Press softkey to select ch.A GROUND. 9.003 EVAL Is only a straight line displayed? 10.001 HEAD {C.1 INPUT COUPLING: ch.B} 10.002 DISP Connect the 5500A Scope Output to ch.B. 10.003 5500 120mVpp 1kH SI SC S L 10.004 DISP Press AUTO. 10.005 DISP Press softkey to select ch.B DC. 10.005 DISP Press softkey to select ch.B 50 Ohm. 10.005 DISP Set VERTICAL AMPL/DIV ch.B to 20mV. 10.005 DISP Set HORIZONTAL TIME/DIV to 20ms. 10.005 DISP Press TRIGGER COUPLING. 10.005 DISP Press softkey twice to select DC. 10.006 MESS Set TRIGGER LEVEL for a triggered display. 10.007 5500 40 240mVpp 10H SI SC N L 10.008 DISP Press VERTICAL COUPLING. 10.008 DISP Press softkey to select ch.B AC. 10.009 EVAL Is the amplitude of the displayed signal >=4.3 div? 11.001 MESS 11.002 DISP Press softkey to select ch.B GROUND. 11.003 EVAL Is only a straight line displayed? 12.001 DISP Press softkey to select ch.A & ch.B 1 MOhm. 12.002 ASK+ R 12.003 HEAD {C.2 VERTICAL DEFLECTION COEFFICIENTS: ch.A} 12.004 MEM2 =ch.A 12.005 DISP Disconnect the 5500A from the UUT. 12.006 DISP Connect the 5500A NORMAL HI and LO to UUT Channel A 12.006 DISP using a dual banana to BNC adapter. 12.007 5500 20mVpp 1kH SI S 2W 12.008 DISP Press AUTO. 12.009 DISP Press softkey to select ch.A DC. 12.010 5500 5 20mVpp 2% 1kH SI 2W 13.001 5500 10 50mVpp 2% 1kH SI 2W 14.001 5500 20 100mVpp 2% 1kH SI 2W 15.001 5500 50 200mVpp 2% 1kH SI 2W 16.001 5500 100 500mVpp 2% 1kH SI 2W 17.001 5500 200 1000mVpp 2% 1kH SI 2W 18.001 5500 500 2000mVpp 2% 1kH SI 2W 19.001 5500 1 5Vpp 2% 1kH SI 2W 20.001 5500 2 10Vpp 2% 1kH SI 2W 21.001 5500 5 20Vpp 2% 1kH SI 2W 22.001 DISP Set VERTICAL VARIABLE ch.A to 10V. 22.002 5500 20Vpp 1kH SI S 2W 22.003 EVAL Is signal amplitude 1.7 to 2.3 divisions? 23.001 DISP Set VERTICAL VARIABLE ch.A to CAL. 23.002 HEAD {C.2 VERTICAL DEFLECTION COEFFICIENTS: ch.B} 23.003 MEM2 =ch.B 23.004 DISP Connect the 5500A NORMAL HI and LO to UUT Channel B 23.004 DISP using a dual banana to BNC adapter. 23.005 5500 20mVpp 1kH SI S 2W 23.006 DISP Press AUTO. 23.007 DISP Press softkey to select ch.B DC. 23.008 5500 5 20mVpp 2% 1kH SI 2W 24.001 5500 10 50mVpp 2% 1kH SI 2W 25.001 5500 20 100mVpp 2% 1kH SI 2W 26.001 5500 50 200mVpp 2% 1kH SI 2W 27.001 5500 100 500mVpp 2% 1kH SI 2W 28.001 5500 200 1000mVpp 2% 1kH SI 2W 29.001 5500 500 2000mVpp 2% 1kH SI 2W 30.001 5500 1 5Vpp 2% 1kH SI 2W 31.001 5500 2 10Vpp 2% 1kH SI 2W 32.001 5500 5 20Vpp 2% 1kH SI 2W 33.001 DISP Set VERTICAL VARIABLE ch.B to 10V. 33.002 5500 20Vpp 1kH SI S 2W 33.003 EVAL Is signal amplitude 1.7 to 2.3 divisions? 34.001 DISP Set VERTICAL VARIABLE ch.B to CAL. 34.002 ASK+ B 34.003 ASK- R 34.004 HEAD {C.3 INPUT IMPEDANCE AND CAPACITANCE: ch.A} 34.005 DISP Connect the 5500A NORMAL HI and LO to ch.A via a 34.005 DISP dual banana adapter and a 1 MOhm/14pF dummy probe 2:1. 34.006 5500 50mVpp 1kH SI S 2W 34.007 DISP Press AUTO. 34.008 DISP Press softkey to select ch.A DC. 34.009 MESS Is signal free from overshoot or undershoot? 34.009 MESS If "YES" press SLEW to check amplitude. 34.010 DISP Set VERTICAL AMPL/DIV ch.A to 5mV. 34.011 5500 10 50mVpp 2% 1kH SI 2W 35.001 DISP Set VERTICAL AMPL/DIV ch.A to 10mV. 35.002 5500 20 100mVpp 2% 1kH SI 2W 36.001 DISP Set VERTICAL AMPL/DIV ch.A to 20mV. 36.002 5500 40 200mVpp 2% 1kH SI 2W 37.001 DISP Set VERTICAL AMPL/DIV ch.A to 50mV. 37.002 5500 100 500mVpp 2% 1kH SI 2W 38.001 DISP Set VERTICAL AMPL/DIV ch.A to 100mV. 38.002 5500 0.2 1Vpp 2% 1kH SI 2W 39.001 DISP Set VERTICAL AMPL/DIV ch.A to 200mV. 39.002 5500 0.4 2Vpp 2% 1kH SI 2W 40.001 DISP Set VERTICAL AMPL/DIV ch.A to 500mV. 40.002 5500 1 5Vpp 2% 1kH SI 2W 41.001 DISP Set VERTICAL AMPL/DIV ch.A to 1V. 41.002 5500 2 10Vpp 2% 1kH SI 2W 42.001 DISP Set VERTICAL AMPL/DIV ch.A to 2V. 42.002 5500 4 20Vpp 2% 1kH SI 2W 43.001 DISP Set VERTICAL AMPL/DIV ch.A to 5V. 43.002 5500 10 50Vpp 2% 1kH SI 2W 44.001 HEAD {C.3 INPUT IMPEDANCE AND CAPACITANCE: ch.B} 44.002 DISP Connect the 5500A NORMAL HI and LO to ch.B via a 44.002 DISP dual banana adapter and a 1 MOhm/14pF dummy probe 2:1. 44.003 5500 50mVpp 1kH SI S 2W 44.004 DISP Press AUTO. 44.005 DISP Press softkey to select ch.B DC. 44.006 DISP Set VERTICAL AMPL/DIV ch.B to 5mV. 44.007 5500 10 50mVpp 2% 1kH SI 2W 45.001 DISP Set VERTICAL AMPL/DIV ch.B to 10mV. 45.002 5500 20 100mVpp 2% 1kH SI 2W 46.001 DISP Set VERTICAL AMPL/DIV ch.B to 20mV. 46.002 5500 40 200mVpp 2% 1kH SI 2W 47.001 DISP Set VERTICAL AMPL/DIV ch.B to 50mV. 47.002 5500 100 500mVpp 2% 1kH SI 2W 48.001 DISP Set VERTICAL AMPL/DIV ch.B to 100mV. 48.002 5500 0.2 1Vpp 2% 1kH SI 2W 49.001 DISP Set VERTICAL AMPL/DIV ch.B to 200mV. 49.002 5500 0.4 2Vpp 2% 1kH SI 2W 50.001 DISP Set VERTICAL AMPL/DIV ch.B to 500mV. 50.002 5500 1 5Vpp 2% 1kH SI 2W 51.001 DISP Set VERTICAL AMPL/DIV ch.B to 1V. 51.002 5500 2 10Vpp 2% 1kH SI 2W 52.001 DISP Set VERTICAL AMPL/DIV ch.B to 2V. 52.002 5500 4 20Vpp 2% 1kH SI 2W 53.001 DISP Set VERTICAL AMPL/DIV ch.B to 5V. 53.002 5500 10 50Vpp 2% 1kH SI 2W 54.001 MESS 54.002 ASK+ R 54.003 ASK- B 54.004 DISP Disconnect the 5500A from the UUT. 54.005 HEAD {C.4 PROTECTION OF 50 Ohm INPUT IMPEDANCE: ch.A} 54.006 DISP Connect the 5500A NORMAL HI and LO to UUT Channel A 54.006 DISP using a dual banana to BNC adapter. 54.007 5500 6Vpp 1kH SQ S 2W 54.008 DISP Press AUTO. 54.009 EVAL Is it possible to set ch.A input impedance to 50 Ohm? 55.001 5500 15Vpp 1kH SQ S 2W 55.002 MESS Did the UUT automatically switch ch.A input impedance 55.003 EVAL back to 1 MOhm? 56.001 MESS 56.002 5500 15Vpp 1kH SQ S 2W 56.003 EVAL Is it possible to set ch.A input impedance to 50 Ohm? [N] 57.001 HEAD {C.4 PROTECTION OF 50 Ohm INPUT IMPEDANCE: ch.B} 57.002 DISP Connect the 5500A NORMAL HI and LO to UUT Channel B 57.002 DISP using a dual banana to BNC adapter. 57.003 5500 6Vpp 1kH SQ S 2W 57.004 DISP Press AUTO. 57.005 EVAL Is it possible to set ch.B input impedance to 50 Ohm? 58.001 5500 15Vpp 1kH SQ S 2W 58.002 MESS Did the UUT automatically switch ch.B input impedance 58.003 EVAL back to 1 MOhm? 59.001 MESS 59.002 5500 15Vpp 1kH SQ S 2W 59.003 EVAL Is it possible to set ch.B input impedance to 50 Ohm? [N] 60.001 DISP Disconnect the 5500A from the UUT. 60.002 HEAD {C.5 VERTICAL DYNAMIC RANGE: ch.A} 60.003 MEM2 =ch.A 60.004 DISP Connect the 5500A SCOPE Output to UUT Channel A. 60.005 5500 2Vpp 200MH LS SC S L 60.006 DISP Press AUTO. 60.007 DISP Press softkey to select ch.A 50 Ohm. 60.008 5500 500 2000mVpp 200MH LS SC N L 60.009 DISP Set VERTICAL AMPL/DIV ch.A to 200mV. 60.010 MESS Shift the displayed sinewave upwards and downwards. 60.011 EVAL Is it free from distortion? 61.001 MESS 61.002 HEAD {C.5 VERTICAL DYNAMIC RANGE: ch.B} 61.003 MEM2 =ch.B 61.004 DISP Connect the 5500A SCOPE Output to UUT Channel B. 61.005 5500 2Vpp 200MH LS SC S L 61.006 DISP Press AUTO. 61.007 DISP Press softkey to select ch.B 50 Ohm. 61.008 5500 500 2000mVpp 200MH LS SC N L 61.009 DISP Set VERTICAL AMPL/DIV ch.B to 200mV. 61.010 MESS Shift the displayed sinewave upwards and downwards. 61.011 EVAL Is it free from distortion? 62.001 MESS 62.002 HEAD {C.6 VERTICAL BANDWIDTH: ch.A 50 Ohm} 62.003 MEM2 =ch.A 62.004 DISP Connect the 5500A SCOPE Output to UUT Channel A. 62.005 5500 120mVpp 100kH SI SC S L 62.006 DISP Press AUTO. 62.007 DISP Press VERTICAL COUPLING. 62.007 DISP Press softkey to select ch.A 50 Ohm. 62.008 5500 20 120mVpp 100kH SI SC N L 62.009 DISP Set HORIZONTAL TIME/DIV to 5ns. 62.010 EVAL Is the displayed amplitude >=4.3 div? 63.001 DISP Disconnect the 5500A from the UUT. 63.002 HEAD {C.6 VERTICAL BANDWIDTH: ch.A 1 MOhm} 63.003 DISP Connect a 50 Ohm terminator directly to the 5500A SCOPE 63.003 DISP Output. 63.003 DISP 63.003 DISP Connect a 10:1 attenuator probe PM8929/09 to ch.A. 63.003 DISP 63.003 DISP NOTE: This probe must be correctly compensated according 63.003 DISP to the procedure given in chapter 8.1.1.4 in the 63.003 DISP operating manual. 63.004 DISP Put the probe tip via a BNC/probe tip adapter in the 63.004 DISP 50 Ohm terminator. 63.005 DISP Set HORIZONTAL TIME/DIV to 2us. 63.006 ASK- R 63.007 5500 200 1200mVpp 100kH LS SC N L 63.008 ASK+ R 63.009 5500 mVpp 200MH LS SC S L 63.010 DISP Set HORIZONTAL TIME/DIV to 5ns. 63.011 EVAL Is the displayed amplitude >=4.3 div? 64.001 DISP Disconnect the probe from the UUT and the terminator 64.001 DISP from the 5500A. 64.002 HEAD {C.6 VERTICAL BANDWIDTH: ch.B 50 Ohm} 64.003 MEM2 =ch.B 64.004 DISP Connect the 5500A SCOPE Output to ch.B. 64.005 5500 120mVpp 100kH LS SC S L 64.006 DISP Press AUTO. 64.007 DISP Press softkey to select ch.B 50 Ohm. 64.008 5500 20 120mVpp 100kH LS SC N L 64.009 5500 mVpp 200MH LS SC S L 64.010 DISP Set HORIZONTAL TIME/DIV to 5ns. 64.011 EVAL Is the displayed amplitude >=4.3 div? 65.001 DISP Disconnect the 5500A from the UUT. 65.002 HEAD {C.6 VERTICAL BANDWIDTH: ch.B 1 MOhm} 65.003 DISP Connect a 50 Ohm terminator directly to the 5500A SCOPE 65.003 DISP Output. 65.003 DISP 65.003 DISP Connect a 10:1 attenuator probe PM8929/09 to ch.B. 65.003 DISP 65.003 DISP NOTE: This probe must be correctly compensated according 65.003 DISP to the procedure given in chapter 8.1.1.4 in the 65.003 DISP operating manual. 65.004 DISP Put the probe tip via a BNC/probe tip adapter in the 65.004 DISP 50 Ohm terminator. 65.005 DISP Set HORIZONTAL TIME/DIV to 2us. 65.006 ASK- R 65.007 5500 200 1200mVpp 100kH LS SC N L 65.008 ASK+ R 65.009 5500 mVpp 200MH LS SC S L 65.010 DISP Set HORIZONTAL TIME/DIV to 5ns. 65.011 EVAL Is the displayed amplitude >=4.3 div? 66.001 DISP Disconnect the probe from the UUT and the terminator 66.001 DISP from the 5500A. 66.002 HEAD {C.7 BANDWIDTH LIMITER: ch.A} 66.003 MEM2 =ch.A 66.004 DISP Connect the 5500A SCOPE Output to ch.A. 66.005 5500 120mVpp 20MH LS SC S L 66.006 DISP Press AUTO. 66.007 DISP Press softkey to select ch.A 50 Ohm. 66.008 5500 20 120mVpp 20MH LS SC N L 66.009 DISP Press VERTICAL MODE. 66.009 DISP Press softkey PROCESSING>. 66.009 DISP Press softkey to select BW LIMIT. 66.010 EVAL Is the displayed amplitude 4.3 div? 67.001 HEAD {C.7 BANDWIDTH LIMITER: ch.B} 67.002 MEM2 =ch.B 67.003 DISP Connect the 5500A SCOPE Output to ch.B. 67.004 5500 120mVpp 20MH LS SC S L 67.005 DISP Press AUTO. 67.006 DISP Press VERTICAL COUPLING. 67.006 DISP Press softkey to select ch.B 50 Ohm. 67.007 5500 20 120mVpp 20MH LS SC N L 67.008 DISP Press VERTICAL MODE. 67.008 DISP Press softkey PROCESSING>. 67.008 DISP Press softkey to select BW LIMIT. 67.009 EVAL Is the displayed amplitude 4.3 div? 68.001 DISP Disconnect the 5500A from the UUT. 68.002 HEAD {C.8 PULSE RESPONSE: ch.A 50 Ohm} 68.003 MEM2 =ch.A 68.004 DISP Connect the 5500A SCOPE Output to ch.A. 68.005 5500 1000mVpp 1MH ED SC S L 68.006 DISP Press AUTO. 68.007 5500 200 1000mVpp 1MH ED SC N L 68.008 DISP Set HORIZONTAL TIME/DIV to 5ns. 68.009 MESS During the first 10ns are the pulse abberations within 68.010 EVAL +/-0.3 div? 69.001 MESS 69.002 DISP Disconnect the 5500A from the UUT. 69.003 HEAD {C.8 PULSE RESPONSE: ch.A 1 MOhm} 69.004 DISP Connect a 50 Ohm terminator to the 5500A SCOPE Output. 69.004 DISP 69.004 DISP Connect a 10:1 attenuator probe PM8929/09 to ch.A. 69.004 DISP 69.004 DISP NOTE: This probe must be correctly compensated according 69.004 DISP to the procedure given in chapter 8.1.1.4 in the 69.004 DISP operating manual. 69.005 DISP Put the probe tip via a BNC/probe tip adapter in the 69.005 DISP 50 Ohm terminator. 69.006 DISP Set HORIZONTAL TIME/DIV to 500ns. 69.006 DISP Set VERTICAL AMPL/DIV ch.A to 200mV. 69.007 5500 200 1000mVpp 1MH ED SC N L 69.008 DISP Set HORIZONTAL TIME/DIV to 5ns. 69.009 MESS During the first 10ns are the pulse abberations within 69.010 EVAL +/-0.3 div? 70.001 MESS 70.002 DISP Disconnect the probe from the UUT and the terminator 70.002 DISP from the 5500A. 70.003 HEAD {C.8 PULSE RESPONSE: ch.B 50 Ohm} 70.004 MEM2 =ch.B 70.005 DISP Connect the 5500A SCOPE Output to ch.B. 70.006 5500 1000mVpp 1MH ED SC S L 70.007 DISP Press AUTO. 70.008 5500 200 1000mVpp 1MH ED SC N L 70.009 DISP Set HORIZONTAL TIME/DIV to 5ns. 70.010 MESS During the first 10ns are the pulse abberations within 70.011 EVAL +/-0.3 div? 71.001 MESS 71.002 DISP Disconnect the 5500A from the UUT. 71.003 HEAD {C.8 PULSE RESPONSE: ch.B 1 MOhm} 71.004 DISP Connect a 50 Ohm terminator to the 5500A SCOPE Output. 71.004 DISP 71.004 DISP Connect a 10:1 attenuator probe PM 8929/09 to ch.B. 71.004 DISP 71.004 DISP NOTE: This probe must be correctly compensated according 71.004 DISP to the procedure given in chapter 8.1.1.4 in the 71.004 DISP operating manual. 71.005 DISP Put the probe tip via a BNC/probe tip adapter in the 71.005 DISP 50 Ohm terminator. 71.006 DISP Set HORIZONTAL TIME/DIV to 500ns. 71.006 DISP Set VERTICAL AMPL/DIV ch.B to 200mV. 71.007 5500 200 1000mVpp 1MH ED SC N L 71.008 DISP Set HORIZONTAL TIME/DIV to 5ns. 71.009 MESS During the first 10ns are the pulse abberations within 71.010 EVAL +/-0.3 div? 72.001 MESS 72.002 DISP Disconnect the probe from the UUT and the terminator 72.002 DISP from the 5500A. 72.003 HEAD {C.9 BASE LINE INSTABILITY} 72.004 DISP Press AUTO. 72.005 DISP Press VERTICAL MODE. 72.005 DISP Press softkey to select A and B. 72.005 DISP Set VERTICAL SHIFT ch.A & ch.B to center traces. 72.006 DISP Press softkey to select ADD. 72.007 EVAL Is the trace centered vertically on the screen +/-0.3 div? 73.001 DISP Press softkey to select A and B. 73.001 DISP Set VERTICAL SHIFT ch.A so trace is 1 div. above mid. 73.001 DISP Set VERTICAL SHIFT ch.B so trace is 1 div. below mid. 73.002 DISP Press softkey PROCESSING>. 73.003 MESS Press softkey repeatedly to select 73.003 MESS MIN / MAXm (on) and MIN / MAX (off). 73.004 EVAL Is the jump of each trace <=0.5 div? 74.001 DISP Press softkey to select MIN / MAX off. 74.002 MESS Switch VERTICAL AMPL/DIV ch.A and ch.B through their 74.002 MESS complete ranges. This can be done by pressing and 74.002 MESS holding down either side of the switch. 74.003 EVAL Is the jump of each trace <=0.15 div? 75.001 DISP Set VERTICAL AMPL/DIV ch.A and ch.B to 200mV. 75.002 DISP Press softkey RETURN. 75.003 MESS Press softkey repeatedly to select 75.003 MESS A-INVERTm (on) and A-INVERT (off). 75.004 EVAL Is the jump of the ch.A trace <=0.3 div? 76.001 DISP Press softkey to select A-INVERT off. 76.002 MESS Press softkey repeatedly to select 76.002 MESS B-INVERTm (on) and B-INVERT (off). 76.003 EVAL Is the jump of the ch.B trace <=0.3 div? 77.001 DISP Press softkey to select B-INVERT off. 77.002 DISP Press VERTICAL COUPLING. 77.003 MESS Press softkey repeatedly to select ch.A AC, DC 77.003 MESS and GROUND. 77.004 EVAL Is the jump of the ch.A trace <=0.1 div? 78.001 DISP Press softkey to select ch.A DC. 78.002 MESS Press softkey repeatedly to select ch.B AC, DC 78.002 MESS and GROUND. 78.003 EVAL Is the jump of the ch.B trace <=0.1 div? 79.001 DISP Press softkey to select ch.B DC. 79.002 DISP Ground ch.A externally with a BNC/banana adapter with 79.002 DISP interconnected outlets. 79.003 MESS Switch VERTICAL VARIABLE ch.A repeatedly between its 79.003 MESS extreme positions. 79.004 EVAL Is the jump of the ch.A trace <=0.6 div? 80.001 DISP Set VERTICAL VARIABLE ch.A to CAL. 80.002 DISP Move the BNC/banana adapter to ch.B. 80.003 MESS Switch VERTICAL VARIABLE ch.B repeatedly between its 80.003 MESS extreme positions. 80.004 EVAL Is the jump of the ch.B trace <=0.6 div? 81.001 DISP Remove the BNC/banana adapter. 81.002 DISP Set VERTICAL VARIABLE ch.B to CAL. 81.002 DISP Set VERTICAL SHIFT ch.A and ch.B to center traces. 81.002 DISP Press MAGNIFY. 81.003 MESS Press softkey to repeatedly select Yx1 81.003 MESS and Yx5. 81.004 EVAL Is the jump of each trace <=0.5 div? 82.001 MESS 82.002 DISP Press softkey to select Yx1. 82.003 HEAD {C.10 COMMON MODE REJECTION RATIO} 82.004 DISP Connect the 5500A SCOPE Output to ch.A and ch.B via 82.004 DISP a 50 Ohm terminator and BNC-T. 82.005 5500 1.6Vpp 100kH SI SC S L 82.006 DISP Press AUTO. 82.007 DISP Press VERTICAL COUPLING. 82.007 DISP Press softkeys to select ch.A & ch.B 1 MOhm. 82.007 DISP Set VERTICAL AMPL/DIV ch.A and ch.B to 200mV. 82.007 DISP Set VERTICAL SHIFT ch.A and ch.B to center traces. 82.008 DISP Press VERTICAL MODE. 82.008 DISP Press softkey to select ADD. 82.008 DISP Press softkey to select B-INVERT on. 82.008 DISP Set VERTICAL VARIABLE ch.A or ch.B for minimal trace 82.008 DISP amplitude. 82.009 5500 1.6Vpp 1MH LS SC S L 82.010 EVAL Is the trace amplitude <=0.08 div? 83.001 5500 1.6Vpp 50MH LS SC S L 83.002 EVAL Is the trace amplitude <=0.4 div? 84.001 DISP Disconnect the 5500A from the UUT. 84.002 HEAD {C.11 AVERAGE FUNCTION} 84.003 OPBR This test requires a another function generator in 84.003 OPBR addition to the 5500A. 84.003 OPBR Do you to perform the test? 84.004 JMPT 84.007 84.005 RSLT =#### Not tested #### 84.006 JMP 85.003 84.007 DISP Press VERTICAL COUPLING. 84.007 DISP Press softkeys to select ch.A & ch.B 50 Ohm. 84.008 DISP Connect the function generator to ch.A. 84.009 DISP Setup the function generator to output a 84.009 DISP 1.2Vpp, 1kHz sinewave. 84.010 DISP Press AUTO. 84.011 DISP Press TRIGGER COUPLING. 84.011 DISP Press softkey to select AC. 84.011 DISP Press softkey to select HF reject on 84.012 DISP Connect the 5500A SCOPE Output to ch.B. 84.013 5500 0.2Vpp 100MH LS SC S L 84.014 DISP Press VERTICAL MODE. 84.014 DISP Press softkey to select ADD. 84.014 DISP Set TRIGGER LEVEL for a triggered display. 84.015 DISP Press softkey PROCESSING>. 84.015 DISP Press softkey AVERAGE >. 84.016 MESS Does the display become cleaner (noise eliminated) 84.016 MESS as averaging is increased sequentially from 84.017 EVAL C = 2 to C = 64? 85.001 MESS 85.002 DISP Disconnect the function generator and 5500A from 85.002 DISP the UUT. 85.003 HEAD {D.1 DEFLECTION COEFFICIENTS TIME BASE} 85.004 DISP Press AUTO. 85.005 DISP Press VERTICAL COUPLING. 85.005 DISP Press softkey to select ch.A DC. 85.005 DISP Press softkey to select ch.A 50 Ohm. 85.005 DISP Set VERTICAL AMPL/DIV ch.A to 500mV. 85.005 DISP Press TRIGGER COUPLING. 85.005 DISP Press softkey to select DC. 85.006 DISP Connect the 5500A Scope Output to ch.A. 85.007 MESS Set TRIGGER LEVEL for a triggered display. 85.008 5500 5 5T 1% MK SC L 86.001 5500 2 2T 1% MK SC L 87.001 5500 1 1T 1% MK SC L 88.001 5500 500 500mT 1% MK SC L 89.001 5500 200 200mT 1% MK SC L 90.001 5500 100 100mT 1% MK SC L 91.001 5500 50 50mT 1% MK SC L 92.001 5500 20 20mT 1% MK SC L 93.001 5500 10 10mT 1% MK SC L 94.001 5500 5 5mT 1% MK SC L 95.001 5500 2 2mT 1% MK SC L 96.001 5500 1 1mT 1% MK SC L 97.001 5500 500 500uT 1% MK SC L 98.001 5500 200 200uT 1% MK SC L 99.001 5500 100 100uT 1% MK SC L 100.001 5500 50 50uT 1% MK SC L 101.001 5500 20 20uT 1% MK SC L 102.001 5500 10 10uT 1% MK SC L 103.001 5500 5 5uT 1% MK SC L 104.001 5500 2 2uT 1% MK SC L 105.001 5500 1 1uT 1% MK SC L 106.001 5500 500 500nT 1% MK SC L 107.001 5500 200 200nT 1% MK SC L 108.001 5500 100 100nT 1% MK SC L 109.001 5500 50 50nT 1% MK SC L 110.001 5500 20 20nT 1% MK SC L 111.001 5500 10 10nT 1% MK SC L 112.001 5500 5 5nT 1% MK SC L 113.001 DISP Disconnect the 5500A from the UUT. 113.002 HEAD {D.2 TIME BASE LINE INSTABILITY} 113.003 DISP Press AUTO. 113.004 MESS Switch HORIZONTAL TIME/DIV through its complete range. 113.005 EVAL Is the vertical jump of the trace <=0.6 div? 114.001 ASK- N 114.002 HEAD {D.3 TRIGGER SENSITIVITY: ch.A} 114.003 DISP Connect the 5500A SCOPE Output to ch.A. 114.004 5500 0.5Vpp 100kH SI SC S L 114.005 DISP Press AUTO. 114.006 5500 0.5Vpp 30MH LS SC S L 114.007 DISP Set HORIZONTAL TIME/DIV to 20ns. 114.008 MESS Decrease the amplitude to 0.5 div (pp). 114.009 5500 0.5Vpp 30MH LS SC N L 114.010 MESS 114.011 EVAL Is the UUT triggered? 115.001 DISP Set HORIZONTAL TIME/DIV to 5ns. 115.002 MESS Decrease the amplitude to 1.0 div (pp). 115.003 5500 0.5Vpp 200MH LS SC N L 115.004 MESS 115.005 EVAL Is the UUT triggered? 116.001 MESS Adjust amplitude for 3 div (pp). 116.002 5500 1Vpp 300MH LS SC N L 116.003 MESS 116.004 EVAL Is the UUT triggered? 117.001 HEAD {D.3 TRIGGER SENSITIVITY: ch.B} 117.002 DISP Connect the 5500A SCOPE Output to ch.B. 117.003 5500 0.5Vpp 100kH LS SC S L 117.004 DISP Press VERTICAL COUPLING. 117.004 DISP Press softkey to select ch.B 50 Ohm. 117.004 DISP Press AUTO. 117.005 5500 0.5Vpp 30MH LS SC S L 117.006 DISP Set HORIZONTAL TIME/DIV to 20ns. 117.007 MESS Decrease the amplitude to 0.5 div (pp). 117.008 5500 0.5Vpp 30MH LS SC N L 117.009 MESS 117.010 EVAL Is the UUT triggered? 118.001 DISP Set HORIZONTAL TIME/DIV to 5ns. 118.002 MESS Decrease the amplitude to 1.0 div (pp). 118.003 5500 0.5Vpp 200MH LS SC N L 118.004 MESS 118.005 EVAL Is the UUT triggered? 119.001 MESS Adjust amplitude for 3 div (pp). 119.002 5500 1Vpp 300MH LS SC N L 119.003 MESS 119.004 EVAL Is the UUT triggered? 120.001 ASK+ N 120.002 HEAD {D.4 TRIGGER SENSITIVITY: EXT} 120.003 DISP Connect the 5500A SCOPE Output to ch.A and EXT TRIG via 120.003 DISP a 50 Ohm terminator, a BNC-T and equal length cables. 120.004 5500 100mVpp 100kH LS SC S L 120.005 DISP Press softkey to select ch.A 1 MOhm. 120.006 DISP Press AUTO. 120.007 DISP Set VERTICAL AMPL/DIV ch.A to 20mV. 120.007 DISP Press TRIGGER SOURCE. 120.007 DISP Press softkey to select EXT. 120.007 DISP Set HORIZONTAL TIME/DIV to 20ns. 120.008 5500 50mVpp 30MH LS SC S L 120.009 EVAL Is the UUT triggered? 121.001 DISP Set HORIZONTAL TIME/DIV to 5ns. 121.002 5500 100mVpp 200MH LS SC S L 121.003 EVAL Is the UUT triggered? 122.001 DISP Remove the BNC-T to disconnect the 5500A from ch.A. 122.002 5500 300mVpp 300MH LS SC S L 122.003 EVAL Is the UUT triggered? 123.001 HEAD {D.4 TRIGGER SENSITIVITY: EXT:10} 123.002 DISP Re-install the BNC-T to Connect the 5500A SCOPE Output 123.002 DISP to ch.A and EXT TRIG. 123.003 5500 1Vpp 100kH SI SC S L 123.004 DISP Press AUTO. 123.005 DISP Set VERTICAL AMPL/DIV ch.A to 200mV. 123.005 DISP Press TRIGGER SOURCE. 123.005 DISP Press softkey to select EXT:10. 123.005 DISP Set HORIZONTAL TIME/DIV to 20ns. 123.006 5500 500mVpp 30MH LS SC S L 123.007 EVAL Is the UUT triggered? 124.001 DISP Set HORIZONTAL TIME/DIV to 5ns. 124.002 5500 1Vpp 200MH LS SC S L 124.003 EVAL Is the UUT triggered? 125.001 DISP Remove the BNC-T to disconnect the 5500A from ch.A. 125.002 5500 3Vpp 300MH LS SC S L 125.003 EVAL Is the UUT triggered? 126.001 DISP Disconnect the 5500A from the UUT. 126.002 HEAD {D.5 LEVEL CONTROL RANGE: ch.A} 126.003 DISP Connect the 5500A Scope Output to ch.A. 126.004 5500 1.6Vpp 50kH SI SC S L 126.005 DISP Press VERTICAL COUPLING. 126.005 DISP Press softkey to select ch.A 50 Ohm. 126.005 DISP Press AUTO. 126.006 DISP Set VERTICAL AMPL/DIV ch.A to 200mV. 126.007 5500 0.2 1.6Vpp 50kH SI SC N L 126.008 DISP Press TRIGGER COUPLING. 126.008 DISP Press softkey to select AC. 126.008 DISP Set VERTICAL AMPL/DIV ch.A to 100mV. 126.009 MESS Does the range of VERTICAL SHIFT ch.A lie at least 126.010 EVAL between top and bottom of the sinewave? 127.001 MESS 127.002 HEAD {D.5 LEVEL CONTROL RANGE: ch.B} 127.003 DISP Connect the 5500A Scope Output to ch.B. 127.004 5500 1.6Vpp 50kH LS SC S L 127.005 DISP Press AUTO. 127.006 DISP Set VERTICAL AMPL/DIV ch.B to 200mV. 127.007 5500 0.2 1.6Vpp 50kH LS SC N L 127.008 DISP Press TRIGGER COUPLING. 127.008 DISP Press softkey to select AC. 127.008 DISP Set VERTICAL AMPL/DIV ch.B to 100mV. 127.009 MESS Does the range of VERTICAL SHIFT ch.B lie at least 127.010 EVAL between top and bottom of the sinewave? 128.001 MESS 128.002 HEAD {D.6 LEVEL CONTROL RANGE: EXT} 128.003 DISP Connect the 5500A Scope Output to ch.A and EXT TRIG 128.003 DISP via a BNC-T & equal length cables. 128.004 5500 1.6Vpp 50kH LS SC S L 128.005 DISP Press VERTICAL COUPLING. 128.005 DISP Press softkey to select ch.A 1 MOhm. 128.005 DISP Press AUTO. 128.006 DISP Set VERTICAL AMPL/DIV ch.A to 200mV. 128.006 DISP Press TRIGGER SOURCE. 128.006 DISP Press softkey to select EXT. 128.006 DISP Press TRIGGER COUPLING. 128.006 DISP Press softkey to select AC. 128.007 5500 200 1600mVpp 50kH LS SC N L 128.008 MESS Does the range of VERTICAL SHIFT ch.A lie at least 128.009 EVAL between top and bottom of the sinewave? 129.001 MESS 129.002 HEAD {D.6 LEVEL CONTROL RANGE: EXT:10} 129.003 5500 16Vpp 50kH SI SC S 129.004 DISP Press AUTO. 129.005 DISP Set VERTICAL AMPL/DIV ch.A to 2V. 129.005 DISP Press TRIGGER SOURCE. 129.005 DISP Press softkey to select EXT:10. 129.005 DISP Press TRIGGER COUPLING. 129.005 DISP Press softkey to select AC. 129.006 5500 2 16Vpp 50kH SI SC N 129.007 MESS Does the range of VERTICAL SHIFT ch.A lie at least 129.008 EVAL between top and bottom of the sinewave? 130.001 MESS 130.002 HEAD {D.7 BANDWIDTH OF TRIGGER FILTERS} 130.003 5500 100mVpp 1kH SI SC S 130.004 DISP Press AUTO. 130.005 DISP Press softkey to select ch.A DC. 130.005 DISP Press TRIGGER SOURCE. 130.005 DISP Press softkey to select EXT. 130.005 DISP Press TRIGGER COUPLING. 130.005 DISP Press softkey to select AC. 130.005 DISP Press softkey to select HF reject on. 130.005 DISP Set TRIGGER LEVEL for a well triggered display. 130.006 5500 mVpp 100kH SI SC S 130.007 DISP Set HORIZONTAL TIME/DIV to 5us. 130.008 EVAL Can the LEVEL control be adjusted for a triggered display? [N] 131.001 5500 200mVpp 100kH SI SC S 131.002 EVAL Can the LEVEL control be adjusted for a triggered display? 132.001 DISP Disconnect the 5500A from the UUT. 132.002 DISP Connect the 5500A NORMAL HI and LO to UUT ch.A using a 132.002 DISP dual banana to BNC adapter. 132.003 5500 100mVpp 5H SI S 2W 132.004 DISP Set HORIZONTAL TIME/DIV to 50ms. 132.005 EVAL Can the LEVEL control be adjusted for a triggered display? [N] 133.001 5500 200mVpp 5H SI S 2W 133.002 EVAL Can the LEVEL control be adjusted for a triggered display? 134.001 DISP Disconnect the 5500A from the UUT. 134.002 HEAD {D.8 HORIZONTAL EXPANSION} 134.003 DISP Connect the 5500A Scope Output to ch.A. 134.004 5500 100uT MK SC S L 134.005 DISP Press AUTO. 134.006 DISP Set HORIZONTAL TIME/DIV to 200us. 134.006 DISP Set DISPLAY X-EXPAND to 100us. 134.007 ASK- R 134.008 5500 100 100uT 5% MK SC L 135.001 ASK+ R 135.002 DISP Disconnect the 5500A from the UUT. 135.003 HEAD {D.9 VERTICAL POSITIONING RANGE: ch.A} 135.004 DISP Connect the 5500A Scope Output to ch.A. 135.005 5500 1.6Vpp 50kH SI SC S L 135.006 DISP Press VERTICAL COUPLING. 135.006 DISP Press softkey to select ch.A 50 Ohm. 135.006 DISP Press AUTO. 135.007 DISP Set VERTICAL AMPL/DIV ch.A to 200mV. 135.008 5500 200 1600mVpp 50kH SI SC N L 135.009 DISP Set VERTICAL SHIFT ch.A fully CW. 135.010 EVAL Does the trace extend into the bottom half of the screen? [N] 136.001 5500 mVpp 50kH SI SC S L 136.002 DISP Set VERTICAL SHIFT ch.A fully CCW. 136.003 EVAL Does the trace extend into the top half of the screen? [N] 137.001 HEAD {D.9 VERTICAL POSITIONING RANGE: ch.B} 137.002 DISP Connect the 5500A Scope Output to ch.B. 137.003 5500 1.6Vpp 50kH SI SC S L 137.004 DISP Press AUTO. 137.005 DISP Set VERTICAL AMPL/DIV ch.B to 200mV. 137.006 5500 200 1600mVpp 50kH SI SC N L 137.007 DISP Set VERTICAL SHIFT ch.B fully CW. 137.008 EVAL Does the trace extend into the bottom half of the screen? [N] 138.001 5500 mVpp 50kH SI SC S L 138.002 DISP Set VERTICAL SHIFT ch.B fully CCW. 138.003 EVAL Does the trace extend into the top half of the screen? [N] 139.001 HEAD {D.9 HORIZONTAL POSITIONING RANGE} 139.002 5500 mVpp 50kH SI SC S L 139.003 DISP Set VERTICAL SHIFT ch.B to center trace. 139.003 DISP Set DISPLAY X-POSITION fully CW. 139.004 EVAL Does the trace extend into the left half of the screen? [N] 140.001 5500 mVpp 50kH SI SC S L 140.002 DISP Set DISPLAY X-POSITION fully CCW. 140.003 EVAL Does the trace extend into the right half of the screen? [N] 141.001 DISP Disconnect the 5500A from the UUT. 141.002 HEAD {E.1 VOLTAGE CURSOR} 141.003 DISP Connect the 5500A Scope Output to ch.A. 141.004 5500 1Vpp 1kH SI SC S 141.005 DISP Press AUTO. 141.006 DISP Press VERTICAL COUPLING. 141.006 DISP Press softkey to select ch.A 1 MOhm. 141.006 DISP Set VERTICAL AMPL/DIV ch.A to 200mV. 141.006 DISP Press LOCK. 141.006 DISP Press CURSOR/ANALYZE. 141.006 DISP Press softkey to select ON +. 141.007 DISP Position the 1st CURSOR in the horizontal mid of the 141.007 DISP top of the waveform. 141.007 DISP Position the 2nd CURSOR in the horizontal mid of the 141.007 DISP bottom of the waveform. 141.008 EVAL Is the voltage cursor read-out 1.00V +/-0.02V? 142.001 HEAD {E.2 TIME CURSOR} 142.002 5500 1mT MK SC S L 142.003 DISP Press WRITE. 142.003 DISP Press AUTO. 142.004 DISP Set VERTICAL AMPL/DIV ch.A to 500mV. 142.004 DISP Set HORIZONTAL TIME/DIV to 1ms. 142.004 DISP Press LOCK. 142.004 DISP Press softkey to select ON +. 142.005 DISP Position the 1st CURSOR and the 2nd CURSOR so that they 142.005 DISP cover a distance of 8 time marker intervals. 142.006 EVAL Is the time cursor read-out 8ms +/-0.0016ms? 143.001 DISP Press WRITE. 143.002 DISP Disconnect the 5500A from the UUT. 143.003 HEAD {F.1 CALIBRATION FREQUENCY} 143.004 DISP Connect the UUT CAL output to ch.A. 143.005 DISP Press AUTO. 143.006 DISP Press LOCK. 143.006 DISP Press softkey to select ON +. 143.007 DISP Position the 1st CURSOR and the 2nd CURSOR so that they 143.007 DISP cover one period of the signal. 143.008 MESS Is the time cursor read-out 500us +/-1.5us 143.009 EVAL with both halfs of the period equal? 144.001 MESS 144.002 DISP Disconnect the UUT CAL output from ch.A. 144.003 DISP Press WRITE. 144.004 HEAD {F.2 CALIBRATION VOLTAGE} 144.005 DISP Connect the UUT CAL output to the 8842A INPUT HI & LO 144.005 DISP via a 50 Ohm terminator. 144.006 8842 250mV 2.5U 2kH 2W 145.001 DISP Disconnect the 8842A from the UUT. 145.002 HEAD {G.1 VERTICAL PLOT OUTPUT VOLTAGE} 145.003 DISP Plug the plot cable into the DIN plot output socket at 145.003 DISP the rear of the UUT. 145.004 DISP Connect the 5500A Scope Output to ch.A and ch.B. 145.005 5500 1.6Vpp 100H SI SC S 145.006 DISP Press AUTO. 145.007 DISP Press VERTICAL COUPLING. 145.007 DISP Press softkey to select ch.A DC. 145.007 DISP Press softkey to select ch.B DC. 145.007 DISP Set VERTICAL AMPL/DIV ch.A to 200mV. 145.007 DISP Set VERTICAL AMPL/DIV ch.B to 200mV. 145.007 DISP Set VERTICAL SHIFT ch.A and ch.B to center traces. 145.008 MESS Increase amplitude so displayed signal shows clipping. 145.009 5500 2.4Vpp 100H SI SC N 145.010 MESS 145.011 DISP Press DISPLAY. 145.011 DISP Press softkey A versus B>. 145.011 DISP Press softkey to select R0 ON. 145.011 DISP Press SAVE/PLOT. 145.011 DISP Press softkey SELECT >. 145.011 DISP Press softkey ANALOG >. 145.011 DISP Press softkey UP or DOWN to select + 20ms/DT. 145.011 DISP Press SAVE/PLOT. 145.012 DISP Connect the 8842A to the vertical plot output; 145.012 DISP Input Hi to the red banana plug and 145.012 DISP Input Lo to the black banana plug. 145.013 MESS Press softkey ANALOG >. 145.013 MESS The oscilloscope starts the plot action. This can be 145.013 MESS seen because the dot moves from left to right on the 145.013 MESS bottom of the screen. 145.013 MESS Note the maximum 8842A reading. 145.014 8842 1000mV S 2W 145.015 DISP Press softkey STOP. 145.016 MEME 145.017 MEMI Enter the maximum 8842A reading observed. 145.018 MATH MEM = MEM * 1E3 145.019 MEMC 1000mV 30U 146.001 HEAD {G.2 HORIZONTAL PLOT OUTPUT VOLTAGE} 146.002 MEME 146.003 5500 Vpp 100H SI SC S 146.004 DISP Connect the 8842A to the horizontal plot output; 146.004 DISP Input Hi to the blue banana plug and 146.004 DISP Input Lo to the black banana plug. 146.005 8842 1000mV S 2W 146.006 MESS 146.007 DISP Press softkey STOP. 146.008 MEMI Enter the maximum 8842A reading observed. 146.009 MATH MEM = MEM * 1E3 146.010 MEMC 1000mV 30U 147.001 HEAD CONCLUSION OF THE PROCEDURE 147.002 END